OVERVIEW
CALsys -80/50 Autocal is a automatic calibration system for RTD / Thermocouple and also for compensation of cold junction. CALsys -80/50 offers the temperature range of -80 to 50°C. It has large tank which is filled with liquid Methanol. The liquid is heated or cooled to the desired temperature. It has better stability & uniformity than dry block bath and as it uses of stirred liquid.
The system consist of Temperature bath and PC Software, which together contribute to the whole cycle of auto calibration process.The system accepts 4 channels, 4 thermocouples or RTDs. The connection for these channels through special type off locking connectors. The channel configuration can be done with LCD display via touch screen keypad.
The thermocouple microvolt & RTD ohm reading for each channel is monitored with CJC compensation. After the calibration PC software generates a report of actual calibrated values for the inputs.
Key Features
- Large Immersion Depths
- High Accuracy
- High Stability and Uniformity
- Wide Temperature Range
- PC Interface
- Simple to use and Cost effective
-
SPECIFICATIONS
Temperature range -80°C to 50°C Accuracy ±0.5°C Stability ±0.04°C at -80°C
±0.07°C at 0°C
±0.05°C at 50°CUniformity ±0.05°C at -80°C
±0.09°C at 0°C
±0.07°C at 50°CVolume 9 ltr Immersion depth 200mm Access opening 120 x 120 mm Method of control Self tunned PID controller Cooling time 60 min (room temperature to -80°C Resolution 0.1°C Display LCD, °C or °F user-selectable Size (H x W x D) 1200 (H) x 630 (W), 500 (D) mm Weight 100kg Power Requirements 230 VAC, 2KW (50Hz) Computer interface RS – 232 Calibration Accredited calibration certificate provided Environmental operating conditions 0°C to 40°C, 0% to 90% RH (non-condensing) Specifications valid in environmental conditions 13°C … 25°C Input (CALsys -80/50 Autocal) Four channels (one master and three test sensors), high quality universal LEMO connector suitable for T/C (J, K, N, T, R, S, B type) and Rtd Software (CALsys -80/50 Autocal) The calibrator will be provided with software for data recording (Manual Mode) and Test Certificate genration in Auto Mode Data logging (CALsys -80/50 Autocal) Data logging facility with logged data export to computer through LAN port (optional USB)
Tempsens Calsys -80/50 Autocal – Liquid Calibration Bath
$0.00 ex GST
Highly Accurate & Automatic Liquid Bath for Industrial/Laboratory field use. CALsys -80/50 Autocal is a automatic calibration system for RTD / Thermocouple and also for compensation of cold junction. CALsys -80/50 offers the temperature range of -80 to 50°C.