OVERVIEW
CALsys -30/110 Autocal is a Highly Stable Furnace for calibrating RTD. It offers a temperature range of -30 to 110°C. This calibrator can be used on-site for high-temperature calibration and applications like aerospace, oil gas petrochemical, pharmaceutical industry, electric power, automotive and material process industry. The comparison volume is a metallic fixed block of special material, which has a diameter of 32 mm and 120mm long. Low-temperature dry block furnace based on thermoelectric cooling circuitry.
The CALsys -30/110 Autocal is an automatic temperature of the calibration system for thermocouples and RTDs. The system consists of Temperature bath PC software, which together contributes to the whole cycle of the auto-calibration process. The system accepts 4 channels, 4 thermocouples, 4 RTDs.The channel configuration can be done with an LCD display via a touch screen keypad. The thermocouple microvolt & RTD ohm reading for each channel is monitored with CJC compensation. After the calibration process complete the PC software generates a report of actual calibrated values for inputs..
Key Features:
- High Accuracy
- High Stability and Uniformity
- Wide Temperature Range
- PC Interfacing
- Simple to Use and Cost Effective
-
SPECIFICATIONS
Temperature range at 25°C -30°C to 110°C Accuracy ±0.5°C Stability ±0.04°C at -30°C
±0.06°C at 0°C
±0.07°C at 110°CRadial uniformity ±0.05°C at -30°C
±0.07°C at 0°C
±0.08°C at 110°CImmersion depth 120 mm Fixed insert dimensions 32 mm Method of control Self tuned PID controller Heating time 10 min Cooling time 25 min (110°C to -30°C) Resolution 0.1 C (0.01°C to -30°C) Display LCD, °C or °F user-selectable Size (H x W x D) 380(H) x 170 (W) x 188 (D) mm Weight 11kg Power requirements 230 VAC, 500W (50 Hz) Computer interface RS – 232 Calibration Accredited calibration certification provided Environmental operating conditions 0°C to 40°C, 0% to 90% RH (non-condensing) Specifications valid in environmental conditions 13°C … 25°C Input (CALsys -30/110 Autocal) Four channels (one master and three test sensors). high quality universal LEMO connector suitable for both T/C (J, K, N, T, R, S, B Type) and Rtd Software (CALsys -30/110 Autocal) The calibrator will be provided with software for data recording (Manual Mode) and Test Certificate genration in Auto Mode Data Logging (CALsys -30/110 Autocal) Data logging facility with logged data export to computer through LAN port (optional USB)
Tempsens Calsys -30/110 Autocal – Automatic Dry Block Calibrator
$0.00 ex GST
CALsys -30/110 Autocal is a Highly Stable Furnace for calibrating RTD. It offers a temperature range of -30 to 110°C.