OVERVIEW
Features
- Automatic (up to 4 points) and manual (up to 5 points) conductivity calibration
- Adjustable reference temperature and temperature coefficient
- Adjustable cell constant and selectable measurement unit
- Automatic / manual temperature compensation and temperature calibration function
- Conductivity and total dissolved solids (TDS) conversion
- Four (4) TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
- Two (2) salinity calibration curves: NaCl, Seawater
- °C & °F temperature readings
- With real-time clock and date/time stamp for GLP compliance
- Auto log data; built-in memory can store up to 1000 data
- Auto stable, auto hold, and real-time measurement modes
- Auto shut-off (programmable: up to 30 minutes)
- Monochrome LCD with white LED backlight: 50 x 50 mm
- Phono jack output for connecting the meter to computer via PC (USB) cable for data collection and software upgrade and to printer via printer (RS232) cable for printout
- Powered by 2 x AA batteries
- Casing is IP67 waterproof/dustproof, shock & scratch resistant, non-slip
- Packaged with electrode and calibration solutions in carrying case
- 3 years meter warranty
Meter Kit
EC220-K
Package Details:
- LAQUAact EC220 Cond/TDS/Res/Sal/Temp meter
- 2 x AA batteries
- Manual and quick guide
- 9383-10D plastic-body, Ti/Pt black conductivity probe (k=1.0) with integrated temperature sensor
- 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm conductivity standard solutions (60ml each)
- Carrying case
Manufactured by HORIBA
-
SPECIFICATIONS
Model
LAQUA EC220
Conductivity/TDS/Res/Sal/Temp (°C/°F)
Conductivity Range …µS/cm to 200.0 mS/cm (k=1.0) Resolution 0.05% full scale Accuracy ±0.6% full scale; ±1.5% full scale > 18.0 mS/cm Reference Temperature 15 to 30 °C (adjustable) Temperature Coefficient 0.00 to 10.00% (adjustable) Cell Constant 0.1, 1.0, 10.0 (adjustable) Calibration Points Up to 4 (Auto) / Up to 5 (Manual) Measurement Unit S/cm, S/m (selectable / auto-ranging) TDS Range …ppm to 100 ppt (TDS factor=0.5) Resolution 0.01 ppm (mg/L) / 0.1 ppt (g/L) Accuracy ±0.1% full scale TDS Curves Linear (0.40 to 1.00), EN27888, 442, NaCl Resistivity Range 0.000 Ω•cm to 20.0 MΩ•cm Resolution 0.05% full scale Accuracy ±0.6% full scale; ±1.5% full scale > 18.0 MΩ•cm Salinity Range 0.0 to 100.0 ppt / 0.00 to 10.00 % Resolution 0.1 ppt / 0.01% Accuracy ±0.2% full scale Salinity Curves NaCl / Seawater Calibration Option Yes Temperature Range -30.0 to 130.0 °C / -22.0 to 266.0 °F Resolution 0.1 °C / °F Accuracy ± 0.5 °C / ± 0.9 °C Calibration Option Yes Memory Data 1000 Auto Log Data Yes Real-time Clock Yes Date & Time Stamp Yes Measurement Modes Auto Stable, Auto Hold, Real Time Auto-Shut Off Yes (Programmable: up to 30 minutes) Electrode Status Yes (On screen display) Diagnostic Messages Yes Software Upgrade Yes via PC (USB) cable PN 3200779639 PC Communication Yes via PC (USB) cable PN 3200779639 Printer Communication Yes via Printer (RS232) cable PN 3200779638 Meter Inputs BNC (for conductivity electrode), phono (for temperature sensor) Meter Output phono (for PC/printer connection) Display Custom LCD with backlight Housing IP67, shock & scratch resistant, non-slip Power Requirement 2 x AA batteries Battery Life > 500 hours Meter Dimensions 160 (L) x 80 (W) x 40.60 (H) mm Meter Weight Approx. 260g (with batteries) / 216g (without batteries)
Horiba EC220-K – EC, TDS, Salinity, Resistivity, Temp, Data Logger Kit
$1,165.00 ex GST
A Portable Meter kit for the measurement of EC/TDS/Salinity/RES/Temp with Data Logging