OVERVIEW
An optional analog module with four analog inputs that support excitation power, scaling, and statistics measurements. These four, single-ended sensor inputs are configured through HOBOlink. A HOBO RX3000 is required for this product.
Features
- 15-bit resolution and high accuracy
- Each module can have its own logging rate
- Input ranges up to 33V
-
SPECIFICATIONS
Input Channels Four, single-ended, in addition to smart sensor data channels Measurement Range and Accuracy 0–25.6 mA DC, ±5 µA ± 0.15% of reading
0–2.5 V DC, ±0.25 mV ±0.2% of reading
0–5 V DC, ±0.25 mV ±0.2% of reading
0–10 V DC, ±0.3 mV ±0.2% of reading
0–20 V DC, ±0.6 mV ±0.2% of reading
0–33 V DC, ±1.20 mV ±0.2% of readingResolution 15 bits Field Wiring Two- or three-wire via screw terminals, 16–24 AWG Minimum/Maximum Input Voltage 0/33 V DC Minimum/Maximum Input Current 0/25.6 mA Minimum Source Impedance for Current Measurement 20 KΩ Excitation Voltage 12 V DC ±5% at 200 mA maximum per module -
TECH NOTES & WIRING
Wiring Diagram – Monitoring Differential Air Pressure with an RX3000
Linked is a diagram that shows how to connect and configure a Veris Differential Air Pressure Sensor (T-VER-PX-Series) with an RX3000 Station.
Wiring Diagram – Monitoring DC Voltage and Current with the RX3000 Station
Linked is a drawing that shows how to monitor DC voltage and/or current using the RX3000 Remote Monitoring Station via the Four Channel Analog Input Module, RXMOD-A1. For DC current monitoring, the T-VER-H971BP-200 DC Current Transducer is needed also.
Wiring Diagram – Monitoring 24VAC On/Off with an RX3000 Station
Linked is a wiring diagram that shows you how to monitor the on/off state of a 24VAC circuit using the RX3000 station. In addition to the RX3000 and the RXMOD-A1 Four-Channel Analog Input module
Wiring Diagram – HOBO RX3000 – Wiring Analog Sensors to the RXMOD-A! Module
Linked is a diagram that shows how to connect sensors to the optional 4-channel RXMOD-A1 analog input module.
Wiring Diagram – HOBO RX3000 – RXMOD-A1 Voltage Input Logic Switch
Linked is a wiring diagram that shows how to monitor switch closures with an HOBO RX3000 via the optional RXMOD-A1 4-channel analog input module.
Using the In-Situ Aqua TROLL 100/200 Conductivity Sensor with the HOBO RX3000 Station
This application note describes how to connect and operate the In-Situ Aqua TROLL 100 or 200 (analog output) with the HOBO RX3000 station.
Using the Gill Instruments WindSonic Opt 3 Wind Speed and Direction Sensor with the HOBO RX3000 Station
This application note describes how to connect and operate the Gill WindSonic sensor (Opt 3 with analog outputs) with the HOBO RX3000 station.
Sample Data – Monitoring Remote Water Level, Rainfall and Specific Conductance in a Drainage Pond
Linked are screen-captures from our HOBOlink web product that displays measurements made with a HOBO RX3000 station that’s monitoring (among other things) rainfall, water level (Stevens vented water level transducer), and specific conductivity (InSitu Aquatroll model 100) in a roadside drainage pond at the Onset facility.
Sample Data – HOBO RX3000 – RXMOD-A1 – Judd Snow Depth from HOBOlink
Linked are screenshots from HOBOlink depicting data recorded by a HOBO RX3000 station that was connected to the Judd Snow-Depth Sensor.
Connecting a Thermocouple to an RX3000 Station
In order to connect a thermocouple to an RX3000 station, you will need to use a third-party thermocouple transmitter. Please see linked drawing for more wiring details.
HOBO – RX3000 4-channel Analog Module Accessory
$322.00 ex GST
An optional analog module with four analog inputs that support excitation power, scaling, and statistics measurements.